For Better Performance Please Use Chrome or Firefox Web Browser

Rapid nonenzymatic monitoring of glucose and fructose using a CuO/multiwalled carbon nanotube nanocomposite-modified glassy carbon electrode

نویسندگان

 
هاجر شکرچی زاده، علی اصغر انصافی، مهدی کدیور

نام ژورنال 

Chinese Journal of Catalysis

خلاصه

A nonenzymatic electrochemical sensor for glucose and fructose was fabricated that contained a glassy carbon electrode modified with a copper oxide (CuO)/multiwalled carbon nanotube (MWCNT) nanocomposite. The electrochemical properties of the CuO/MWCNT-modified glassy carbon electrode were investigated. Two distinguishable anodic peaks were observed around 0.30 and 0.44 V corresponding to the oxidation of glucose and fructose, respectively, at the surface of the modified electrode. The detection limits for glucose and fructose were both 0.04 mmol/L. The sensor was used to simultaneously determine the concentrations of glucose and fructose in hydrolyzed sucrose samples, and to measure glucose in blood serum samples, demonstrating its potential as a nonenzymatic carbohydrate sensor.

 

Journal Papers
ماه: 
June
سال: 
2013

تحت نظارت وف ایرانی

Rapid nonenzymatic monitoring of glucose and fructose using a CuO/multiwalled carbon nanotube nanocomposite-modified glassy carbon electrode | دکتر هاجر شکرچی زاده

خطا

The website encountered an unexpected error. Please try again later.

تحت نظارت وف ایرانی